2 complete a reading in the multi wavelength mode – Hach-Lange DR 6000 User Manual User Manual

Page 105

Advertising
background image

105

Expanded programs

7. Press OK to confirm.

Concentration Resolution:

1. Press Concentration Resolution in the "Options" menu.

2. Select the resolution.

3. Press OK to confirm.

4. Press Return to return to the results screen.

Reading Mode

The Reading Mode functions enable the sample to be analyzed
with a single reading or with an optional carousel insert (sample
changer).

Single Reading: A single reading is taken after Read is
pressed.

Carousel 1 inch square: Performs a series of
measurements in 1-inch square cells with an option carousel
insert.

Carousel 1 cm square: Performs a series of measurements
in 1-cm square cells with an optional carousel insert.

Additional information can be found in the carousel user manual.
(

section 5.4.9, page 61

).

6.4.2

Complete a reading in the Multi Wavelength mode

1. Insert the zero solution cell into the cell compartment.

2. Press Zero.

Note: The Read key does not become active until the zero reading
has been completed.

3. Insert the analysis cell into the cell compartment.

4. Press Read.

5. For data storage, refer to

section 5.3.1, page 42

.

6.5

Wavelength Scan mode – recording of absorbance and transmission
spectrums

In the Wavelength Scan mode, the absorbance of the light in a
solution over a defined wavelength spectrum is measured.

The reading results can be displayed as a curve, as percentage
transmittance (%T) or as Absorbance (Abs). The collected data
can be printed as a table or a curve.

The data is available for formatting changes. These include
automatic scaling and zoom functions. Maximum and minimum
values are determined and shown as a table.

The cursor can be moved to any point on the curve for the
purpose of reading off the absorbance or transmittance value and