Expanded programs – Hach-Lange DR 6000 User Manual User Manual
Page 104

104
Expanded programs
The following formulas are available:
K
1
A
1
+ K
2
A
2
K
1
A
1
+ K
2
A
2
+ K
3
A
3
K
1
A
1
+ K
2
A
2
+ K
3
A
3
+ K
4
A
4
K
1
A
1
/ K
2
A
2
(K
1
A
1
+ K
2
A
2
) / K
3
A
3
(K
1
A
1
+ K
2
A
2
) / (K
3
A
3
+ K
4
A
4
)
A
1
refers to the absorbance at wavelength 1
A
2
refers to the absorbance at wavelength 2, etc.
K
1
refers to the coefficient at wavelength 1
K
2
refers to the coefficient at wavelength 2, etc.
If a subtraction has to be completed, the factors can be entered
with a minus sign.
6. To change the wavelength, press the
λx key.
7. Enter the required wavelength factor using the alphanumeric
keypad.
8. Press OK to confirm the selection.
9. To change a coefficient, press one of the "K
X
" keys.
10. Enter the required coefficient using the alphanumeric
keypad.
11. Press OK to confirm the selection.
Note: Up to 5-digit numeric values can be entered. A maximum of
four decimals places are possible.
Concentration Factor:
1. Press Concentration Factor: Off in the "Options" menu.
2. Press On to highlight this feature.
3. Press Factor.
4. Enter the factor by which the absorbance values are to be
multiplied.
5. Press Unit.
6. Select the respective concentration unit from the displayed
list.