3 program a new test, Standard programs – Hach-Lange DR 6000 User Manual User Manual

Page 67

Advertising
background image

67

Standard programs

1. Press Stored Programs in the Main Menu and select a

program.

2. Press Program Options and then Edit.

The Data list of the corresponding test will appear.

The process specification of the corresponding test contains
the new data (Wavelength, Factors, reading Ranges,
Conversion Factor for chemical Form etc.).

3. Highlight the corresponding line in the data overview and

press Edit.

4. Press OK and then Store after editing the test.

5.4.12.3 Program a new test

1. Press Stored Programs in the Main Menu.

2. Press Program Options and then New.

3. Use the alphanumeric keypad to enter the Program Number

indicated in the working procedure.

An overview of the data for the test to be programmed is
displayed.

The process specification of the corresponding test contains
the new data (Wavelength, Factors, reading Ranges,
Conversion Factor for chemical Form etc.).

4. Highlight the line containing the test specification that is to be

programmed, then press Edit and enter the data in the
appropriate program lines.

5. Press OK and then Store after editing the test.

5.4.13 Add stored programs to the favorite programs list

The "Favorites" menu simplifies test selection by creating a list of
the most frequently used tests from the Stored Programs and
User Programs.