3 manual check/revision of test data, 4 update via the internet, Standard programs – Hach-Lange DR 6000 User Manual User Manual

Page 76

Advertising
background image

76

Standard programs

5.5.5.3

Manual check/revision of test data

If an incorrect entry was made during the data update and is not
recognized via the barcode of the cuvette/sample cell (e. g. an
incorrect measuring range), another option is available for
checking and if necessary revising, the test data.

1. The device must show the "main menu".

Alternatively, select in the main menu the menu option
Barcode Programs.

2. Insert the zero cell or sample cell (corresponding to the work

instruction) into the cell compartment (1).

3. Press Option\s > More > Edit.

A data overview is displayed, with the specifications of the
test that is to be revised.

4. Compare the displayed data with the data in the process

specification and edit the displayed data in accordance with
the process specification.

5.5.5.4

Update via the Internet

Update via USB memory stick

To obtain the software for the update from the Internet at
www.hach-lange.com:

1. Go to http://www.hach-lange.com.

2. Select the country and go to Download\>Software.

3. Enter DR 6000 in the Search field.

4. Navigate to the corresponding file for download.

5. Save the file to a USB storage device or PC.

6. Press Instrument Update in the "System Check" menu.

7. Connect the USB memory stick to the USB port (type A) on

the instrument, refer to

section 3.5, page 18

.

8. Press OK. The link is established automatically and the

software is updated.

9. Press OK to return to the "System Check" menu.

Update via the web function

1. Connect the instrument to the network via an Ethernet cable.

2. Activate the Internet connection by selecting

Instrument Setup > PC & Printer (

section 5.2.7.4,

page 37

).

Advertising