6 pipette check, Expanded programs – Hach-Lange DR 6000 User Manual User Manual

Page 127

Advertising
background image

127

Expanded programs

1. Select the required AQA measure – for example a standard

measurement here.

2. Perform the standard measurement via stored program or

barcode program.

3. An AQA symbol is displayed in the measurement window,

the AQA key in the toolbar is displayed as yellow.

4. The AQA measure is shown in a log as Passed or Failed.

6.7.4.6

Pipette check

This function and the optional accessories tests the accuracy of
the pipette test set. To carry it out follow the instructions in the
pipette test set.

1. Press Pipette Check.

2. Press Options.

3. Select New.

4. Define the pipette to be checked, the wavelength and the

upper and lower tolerance limits.

5. Select Close.

6. Select the pipette to be checked from the selection list and

press Start.