5 program a new test, Standard programs – Hach-Lange DR 6000 User Manual User Manual
Page 77

77
Standard programs
3. Press Web Function on the toolbar.
4. Press Instrument Update.
The update will start automatically.
Note: When the instrument software has been updated, a prompt to
restart the instrument is displayed.
5.5.5.5
Program a new test
1. The instrument needs to be at the "Main Menu" or press
Barcode Programs in the Main Menu.
2. Insert the zero cell or sample cell (corresponding to the
process specification) into the cell compartment.
The message "Program not available" is shown, followed by
a data overview with the specifications of the test that is to be
revised.
The process specification of the corresponding test contains
the new data (Wavelength, Factors, Reading Ranges,
Conversion Factor for chemical form etc.).
3. Highlight the corresponding line including the specifications
of the test that is to be programmed and press Edit.
4. To finish programming, press Store and then Cancel to
return to the Main Menu.
5. Press Start to begin the test with the new data.
Note: Once the instrument has an active network connection, new
software versions are announced as they become available. An
instrument update can then be performed.