4 delete stored data from the data log, Standard programs – Hach-Lange DR 6000 User Manual User Manual

Page 45

Advertising
background image

45

Standard programs

1. Plug the USB storage device into the USB A interface on the

DR 6000 or connect the DR 6000 with a network drive (refer
to

section 5.2.7.4, page 37

).

2. Press Recall Data from the Main Menu.

3. Select the data category to be transferred, e.g. Data Log.

A list of the selected measurement data is displayed.

4. Tap on Options and then on the

PC & Printer symbol.

5. Select the data to be sent.

The following options are available:

Single point: only this selected reading will be sent

Filtered data: only readings that correspond to the set
filters will be sent

All data: all data in the selected data category will be
sent.

6. Press OK to confirm.

Note: The number in parentheses is the total number of data sets
assigned to this selection.

5.3.1.4

Delete stored data from the data log

1. Press Recall Data from the Main Menu.

2. Press Data Log > Options > Delete.

3. Highlight Single point, Filtered data or All data and press

OK to confirm.

Note: The number in parentheses is the total number of data sets
assigned to this selection.