65

Standard programs

Description of the table of readings

The first column shows the standard addition volume. 0 mL
stands for the sample without added standard.

The second column shows the reading of the samples with
and without added standard.

The third column shows the detection rate of the added
standard.

Note: The blue highlighted line is active.

The measurement result of the sample without added
standard is automatically shown under 0 mL.

9. Prepare the standard addition solution as described in the

procedure.

10. Press the standard addition volume in the table.

11. Insert the cell with the respective addition volume into the

cell compartment.

12. Press Read.

Repeat the procedure with all the other standard addition
solutions.

13. After all the standard addition solutions have been

measured, press Graph.

The regression line through the standard addition data points
is displayed.

The correlation coefficient r

2

indicates how close the data

points are to the line.

If the correlation coefficient = 1, the curve is linear.

The concentration shown above the curve is the estimated
concentration of the sample without the added standard.

Note: In the Graph menu, the name on the Graph key switches to
Table. Press Table to display all the data in the table again.

14. Press Ideal Line to display the relationship between the

additions and the ideal line (detection rate 100 %).