3 send data from the data log, Standard programs – Hach-Lange DR 6000 User Manual User Manual

Page 44

Advertising
background image

44

Standard programs

3. Press Filter: On/Off.

The function Filter Settings is used to search for specific
items.

4. Activate On. The data can now be filtered using the following

selection criteria.

Sample ID

Operator ID

Start Date

Parameter

or any combination of the four.

5. Press OK to confirm the selection.

The chosen items are listed.

6. Press View Details to get more information.

5.3.1.3

Send data from the data log

Data is sent from the internal data log as an XML (Extensible
Markup Language) file or as a CSV (Comma Separated Value)
file, to a directory with the name DataLog on a USB mass storage
device or a network drive. The file can then be processed using a
spreadsheet program. The file name has the format:
DLYear-Month-Day_Hour_Minute_Second.csv or
DLYear-Month-Day_Hour_Minute_Second.xml.

To send data to a printer, refer to

section 5.2.7.2, page 37

.

Advertising