Standard programs
4. Tap on Table to display details.
Note: Press View List to return to the "Recall Data" list.
5.3.3.3
Send data from wavelength scan or time course
There are two ways to send data to a USB memory stick, printer,
PC or installed network drive with Hach Data Trans.
Option 1:
1. Press Recall Data in the Main Menu and then Wavelength
Scan or Time Course.
2. Press Option and then press the PC & printer symbol to
send the data to a USB storage device, printer, PC or
network drive with Hach Data Trans.
•
When a printer is connected, you must select how the
data should be sent to the printer (as graph, table or
graph and table).
•
If a USB storage device is connected, the files are
automatically sent as xml- or csv files to the file folder
"WLData" (data of a wavelength scan) or "TCData" (data
of a time course) on the USB storage device/or moved
into the folder.
•
If a network drive has been installed under Instrument
Setup > PC & Printer > Network > Setup and csv or xml
has been selected as the data format, the data is sent to
the network drive.
The file name will be formatted as: "ScanData_X.csv"
(Wavelength Scan Data) or "TCData_X.csv" (Time Course
Data).
X = number of scans (1–20).
For further processing use a spreadsheet program.
Note: The message, "Data already present. Overwrite?" is displayed if
the files are already stored. Press OK to overwrite the stored data.