106

Expanded programs

the wavelength. The data associated with each data point can
also be shown as a table.

6.5.1

Set up the wavelength scan

Press Wavelength Scan in the Main Menu. Press Options for
Parameter Setup.

Table 14 Options during wavelength scan

Option

Description

More

Further options are displayed

Folder icon

To store the scan data

Reference Off/On

From the displayed list of stored scans, a record is selected for use as a reference
scan/superimposed scan. This can be highlighted or shown in the background in comparison
with the actual measured scan.
Note: This option is only available when there are stored scans with the same wavelength range
and step.

λ

Enter the wavelength spectrum and scan interval

Timer icon

This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.

View
Table/Graph

Enables the switching between the display of tabular scan data (wavelength/absorbance) and
the graphical representation of the curve.
Note: "View Table" is activated after the first measurement.

Cursor Mode

To select Track or Peak/Valley. The selection for this menu option determines to which points
on the graph the cursor moves.

Send Data

To send data to a printer, computer or USB memory stick (Type A)

Integral: On/Off

The integral gives the area and the derivative of the integral gives the original function

Scale & Units

Scale: In the automatic scaling mode, the y-axis is automatically adapted so that the total scan
is shown.
The manual Scaling mode allows sections of the scan to be displayed.
Units: Choice of absorbance or transmittance.

Reading Mode

Single Reading: One measurement result is shown after tapping on Read.
Carousel 1 inch square: Optional measurement of carousel insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel insert with up to 7 square cells.

Recall measurement
data

Recall saved measurement data, wave-length scans or time courses, see

section 5.3, page 42

.

Instrument Setup mode

Basic data of the instrument, refer to

section 5.2, page 25

.