3 main menu, Standard programs 5.1.3 main menu – Hach-Lange DR 6000 User Manual User Manual

Page 24

Advertising
background image

24

Standard programs

5.1.3

Main menu

A variety of modes may be selected from the Main Menu. The
following table briefly describes each menu option.

You find a toolbar on the right in the display. Press to activate the
various functions.

Table 2 Main Menu options

Option

Function

Stored Programs /
Barcode Programs
(HACH-LANGE programs)

Stored programs are pre-programmed methods that make use of HACH chemicals and
HACH-LANGE pipette tests.
The working procedures for HACH-LANGE tests are included in the test packs.
Further information, as well as illustrated, step-by-step process instructions for analyses using
HACH programs, are available on the website of the manufacturer.

User Programs

User programs make "made to measure analysis" possible:

Users can program methods they have developed themselves

Existing HACH- and HACH-LANGE procedures can be stored as user programs.
These procedures can then be modified according to separate requirements.

Favorites

List of methods/tests created by the user to suit his own requirements.

Single Wavelength

Single wavelength readings are:
Absorbance readings: The light absorbed by the sample is measured in absorbance units.
Transmittance reading (%): Measures the percent of the original light that passes through the
sample and reaches the detector.
Concentration readings: A concentration factor can be entered to enable the measured
absorbance values to be converted into concentration values.

Multi Wavelength

In the Multi Wavelength mode, absorbance (Abs) or percentage transmittance (%T) is
measured at up to four wavelengths and absorbance differences and absorbance relationships
are calculated. Simple conversions into concentrations can also be carried out.

Wavelength Scan

A wavelength scan shows how the light from a sample is absorbed over a defined wavelength
spectrum. This function can be used to determine the wavelength at which the maximum
absorbance value can be measured. The absorbance behavior is displayed graphically during
the scan.

Time course

The time scan records the absorbance or % transmittance at a wavelength over a defined
time.

System checks

The "System check" menu offers a number of options, including optical checks, output checks,
lamp history, instrument update, service time, settings for analytical quality assurance and
instrument backup.

Recall measurement data

Stored data can be called up, filtered, sent and deleted.

Instrument Setup

In this mode, user-specific or method-specific settings can be entered: operator ID, sample ID,
date & time, sound, PC & printer, password, energy-saving mode and stored data.

Advertising