Standard programs

Option 2:

1. Press Wavelength Scan or Time Course and then Options

> More > Send Data to send the data to a USB memory
stick or to a printer.

When a printer is connected, select how to send the data
to the printer (graph, table or both graph and table).

If a USB storage device is connected, the files are
automatically sent as xml- or csv files to the file folder
"WLData" (data of a wavelength scan) or "TCData" (data
of a time course) on the USB storage device.

If a network drive has been installed under Instrument
Setup > PC & Printer > Network > Setup
and csv or xml
has been selected as the data format, the data is sent to
the network drive.

The file name will be formatted as: "ScanData_X.csv"
(Wavelength Scan Data) or "TCData_X.csv" (Time Course
Data).
X = number of scans (1–20).

For further processing use a spreadsheet program.

5.3.3.4

Delete stored data from wavelength scan or time course

1. Press Recall Data in the Main Menu and then Wavelength

Scan or Time Course.

A listing of the stored data is displayed.

2. Highlight any data to delete.

3. Press Delete in the "Options" menu and press OK to

confirm.

5.3.4

Data analysis

Consistent use of the sample ID function is necessary for all data
analysis functions. Only data with identical sample IDs can be
used in data analysis functions.

Use the Trends function to display a time-variation curve for the
stored readings for each parameter and location. The
concentration of the the corresponding parameter is shown
graphically over time.

Use the Ratio function to monitor the parameter relationships at a
particular location and display them in a graph.

Use the Interference Check function to display analyses with the
same sample IDs and unfavourable levels of ion interference.