1 verification kit, Expanded programs – Hach-Lange DR 6000 User Manual User Manual

Page 117

Advertising
background image

117

Expanded programs

6.7.3.1

Verification kit

The Verification Kit (refer to

Section 9, Page 145

) is designed for

periodic monitoring of scattered light, photometric accuracy and
the wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.

1. Press Verification Kit in the "Optical Checks" menu.

2. Press Nominal Values.

3. Press Edit.

An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the
quality control certificate, to the following specifications:

Stray light

Photometrical accuracy

Wavelength accuracy

4. Press OK when all values are entered and the overview is

displayed.

5. Press Verification.

6. Insert the universal cell adapter(

Figure 3, Page 19

) into the

cell compartment (2).

7. Remove any cells from the cell compartment.

8. Close the cell compartment cover.

9. Press Start.