128

Expanded programs

7. Insert the zero cell and press Zero.

A zero measurement is performed.

8. Insert the analysis cell and press Read.

A measurement is performed.

A test log indicating the target and actual value and whether
or not the test has been passed is shown.

9. Press the PC & Printer icon to send the test log to the printer.

10. Press Close to return to the selection list.

11. Press Options to set further functions:

New: Defines an additional pipette.

Delete: Deletes a defined pipette.

Edit: Modifies settings for a defined pipette.

Reminder: Defines a reminder interval to determine when
the user is prompted to return to this AQA measurement.

History: List of all previously performed pipette checks.

12. Press Close to exit the options menu.

6.7.5

Instrument backup

Prior to the next service date, the "Instrument Backup" offers
the opportunity to store all programs, measurement data, user
IDs, passwords, as well as all adjustable data on a USB stick.

1. Press Instrument Backup in the "System Check" menu.

2. Insert the USB stick (refer to

section 3.5, page 18

).

3. Press Store to start a Backup.