Section 1 technical data – Hach-Lange DR 6000 User Manual User Manual

Page 9

Advertising
background image

9

Section 1

Technical data

Changes reserved!

Performance specifications

Operating mode

Transmittance (%), Absorbance and Concentration

Source lamp

Deuterium lamp (UV) and halogen lamp (visual range)

Wavelength range

190–1100 nm

Wavelength Accuracy

± 1 nm (wavelength range 200–900 nm)

Wavelength reproducibility

< 0.1 nm

Wavelength resolution

0.1 nm

Wavelength calibration

Automatic

Wavelength selection

Automatic, based on method selection

Scanning speed

900 nm/min (in 1-nm steps)

Spectral bandwidth

2 nm (1.5–2.9 nm at 656 nm, 1 nm for D2 line)

Photometric measuring range

± 3 Abs (wavelength range 200–900 nm)

Photometric accuracy

5 mAbs at 0.0–0.5 Abs
< 1 % at 0.50–2.0 Abs at 546 nm

Photometric linearity

< 0.5 % to 2 Abs
≤ 1 % at > 2 Abs with neutral glass at 546 nm

Stray light

KI solution at 220 nm
< 3.3 Abs / < 0.05 %

Photometric drift against blank
reading
(30-min baseline stable)

190–199 nm +/- 0.0100 Abs
200–349 nm +/- 0.0054 Abs
350–899 nm +/- 0.0034 Abs
900–1100 nm +/- 0.0100 Abs

Long-term stability

Zero point at 546 nm for 10 hours

≤ 0.0034 Abs

Data log

5000 measurement values (result, date, time, sample ID, user ID)
50 scans, 50 time scans

User programs

200

Physical and environmental specifications

Width

500 mm (19.69 in)

Height

215 mm (8.46 in)

Depth

460 mm (18.11 in)

Weight

11 kg (24.25 lb)

Ambient operating requirements

10–40 °C (50–104 °F), maximum 80 % relative humidity
(without condensate formation)

Ambient storage requirements

–25–60 °C (–13–140 °F), maximum 80 % relative humidity
(without condensate formation)

Additional technical data

Power connection

100–240 V/50–60 Hz

Power consumption

150 VA

Fuse

T 2A H; 250 V (2 units)

Advertising