Luminex xPONENT for MAGPIX 4.2 User Manual

Page 110

Advertising
background image

Count - Type the desired bead count for the analytes by clicking in the Count box. If each

selected bead set does not acquire this number of events, a warning is added to the log
that not enough bead events were acquired. If you select bead sets that are not present,
the analyzer continues to acquire, trying to reach the number of events per bead for bead
sets that are not in the sample. Therefore, choose only the bead sets present in your
sample.

Apply All - Applies the information in the Units and Counts boxes to all analytes.

Total Count - Select Stop after bead count reaches: to stop acquisition when the bead

count reaches a certain number determined by the user. Type the desired value in the box.
The default value is 50. You can also specify the minimum allowable bead count per well
that the xPONENT

®

software analyzes. This excludes data from any beads carried over

during acquisition.

Protocols Page

97