Revive after storage routine – Luminex xPONENT for MAGPIX 4.2 User Manual

Page 24

Advertising
background image

NOTE: When you adjust and save the probe height settings for all three

areas under a plate name, all areas retain the adjustment.

WARNING:

Correct sample probe height is critical to successful sample
acquisition and calibration. Problems with the sample probe can
lead to fluid leaks and inhibit sample acquisition.

CAUTION:

Ensure that the probe height is set correctly before calibrating
the system.

FIGURE 1.

Sample Probe Height Adjustment

Revive After Storage Routine

NOTE: The Revive After Storage routine is necessary when the system

runs for the first time and is recommended when the system has
been idle for more than a week.

After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.

1. Open the Maintenance page, then the Cmds & Routines tab.

2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After

Storage routine performs the following commands:

Prime

Rinse

Alcohol Flush

Rinse

Introduction

11