Defining the system initialization routine – Luminex xPONENT for MAGPIX 4.2 User Manual

Page 140

Advertising
background image

NOTE: See "Daily Activities" on the Home page for daily maintenance

routines.

Luminex recommends that you verify daily and calibrate weekly. You should also verify and
calibrate if any of the following occurs:

• The delta calibration temperature exceeds ± 5° C.

• You move the instrument.

• You experience sample acquisition problems.

• You perform maintenance on the instrument, for example, replacing a hardware

component.

Available system initialization procedures are as follows:

• Fluidics preparation, calibration, performance verification

• Fluidics preparation, performance verification

• Fluidics preparation

Defining the System Initialization Routine

NOTE: Luminex recommends daily performance verification and weekly

calibration of the MAGPIX system. You can set up the system
initialization routine to include calibration and verification on the
Admin page, System Setup tab, Maintenance Options section.

1. Open the Admin page.

2. Click System Setup.

Admin Page

127