Luminex xPONENT for MAGPIX 4.2 User Manual

Page 152

Advertising
background image

This tab contains the following:

Alert Options - Offers three check boxes for each alert - Dialog, Email, and Sound. This
section displays the following events:

Locked User - The user is locked out of xPONENT.

Batch Complete - The batch is finished.

Heater In Range - The plate heater is warmed to the designated range.

Scheduled Maintenance - A scheduled maintenance routine is now due.

Low Bead Count Detected - A bead count below the accepted range is encountered.

System Alerts - One of the following system events has occurred:

• Empty Sheath fluid

• Low Sheath fluid

• Full waste container

• Activated bubble sense

• Overloaded syringe pump plunger

• Probe step loss

• Sheath pressure error

• X step loss

• Y step loss

Routine Complete - The currently running routine is complete.

Delta Cal Temp Exceeded Tolerance During Batch - The delta cal temperature fell

outside the designated range during the running of the batch.

Select the appropriate check boxes to display a pop-up message, send an email, or play a
sound to alert you when the event occurs. All check boxes are selected by default. Cancel
and Save buttons display if you change a setting.

Speakers Attached/Speakers Not Attached - Enables sounds to be played through
speakers. Click this button if you have speakers attached to your system. This is a toggle
button.

Setup Email - Opens the Setup Email dialog box.

Enable or disable email notification by selecting or clearing the Email Active check box.

Contact your network administrator to determine the information needed in the Mail Server
Host
, From Email Address, From Email Password, and Mail Server Port boxes, and if
you need to select the Enable SSL (Secure Sockets Layer) check box.

Admin Page

139