Analytes grid - A grid representing each analyte from 12 to 78. Select All selects all

analytes, and Deselect All deselects all analytes. Click a numbered analyte to select it;
click the analyte again to deselect it. You can also click and drag to select groups of
analytes. Selected analytes are red. Deselected analytes are gray. An analyte marked as
an intra-well normalization bead is blue.

Default Analysis - The default analysis changes based on the Analysis Type selected in

the Settings tab. You can change the analysis settings for all analytes by clicking Change
if this button is enabled in this tab. This displays the Analysis Settings dialog box.

FIGURE 3.

Analysis Settings Dialog Box

If you selected Quantitative on the Settings tab, the default analysis formula is 5P
Weighted
. To change the default, select one of the following from the Method list:

No Analysis

Cubic Spline

Linear Fit

Logistic 4P

Logistic 5P

If you selected Logistic 4P or Logistic 5P, select a weight type of either None or 1/y

2

.

If you selected Qualitative on the Settings tab, the default analysis is Luminex
Qualitative
. Change the default value by selecting either Luminex Qualitative or No
Analysis
.

Click Apply to All Analytes to apply your selection to all selected analytes. Click OK to
change the default analysis to the analysis you selected. Click Cancel to close the dialog
box without saving.

Units - Type the desired units for the analytes in this box.

Batches Page

43