Replaying or recalculating a large batch can take 1 hour or more to complete. Batch replay
cannot be stopped while in progress. Allow adequate time for the operation to complete. The
operation is complete when all progress bars have disappeared.

A batch can be reprocessed multiple times. If the system crashes but the plate finished, the
data can be recovered by replaying the batch.

The initial batch data and output file always remain intact and unchanged. Each time you
replay or recalculate a batch, the system handles it as if it is a new batch, creating a separate
batch entry and output file.

If you select to replay or recalculate a batch that was originally run without a saved protocol,
you have to modify the settings on the following subtabs:

Settings

Analytes

Stds & Ctrls

Plate Layout

These subtabs appear under the Saved Batches tab. After you have completed them in
order, click Replay Batch on the Plate Layout subtab to perform the replay or recalculate
procedure.

Selecting Replay Mode

1. Open the Results page, then the Saved Batches tab.

2. Select the batch that you want to replay and click Replay at the bottom of the screen.

This opens the Select Replay Mode dialog box. By default, Recalculate data is
selected.

3. Select Replay batch or Recalculate data.

The information that you must enter to replay or recalculate is the same.

4. Select the correct protocol and click Next.

5. Select the wells to be acquired and click Replay Batch.

Results Subtab

Results > Saved Batches > Results

xPONENT for MAGPIX 4.2 Software User Manual