Adjusting the sample probe height – Luminex xPONENT for MAGPIX 4.2 User Manual

Page 23

Advertising
background image

1. Adjusting the Sample Probe Height

2. Revive After Storage (Luminex) Routine

3. Calibration/Verification

Adjusting the Sample Probe Height

Adjust the sample probe height to ensure that the probe drops far enough into the well to
acquire sample.

NOTE: Ensure that there is no liquid in the wells or reservoirs before

adjusting the sample probe height.

1. On the Home page, click Probe and Heater under Daily Activities. The Probe &

Heater tab opens.

2. Use well D6 (this is the center of a standard 96-well plate).

3. Ensure that the well location is selected on the plate image. A green pin marks the

selected well.

4. Based on the type of plate you are using, place alignment disks or an alignment sphere in

the well.

• For a standard 96-well plate - none

• For a Filter-bottom plate - two 5.08 mm disks

• For a Mylar-bottom plate - two 5.08 mm disks

• For a conical (v-shaped) plate - one sphere

5. Click Eject to eject the plate carrier.

6. Place the off plate reagent block on the plate carrier. Make sure it is well seated so that it

clips into place.

7. Place a strip well (provided with the Calibration and the Performance Verification kit) in

the off-plate reagent block.

8. In the Strip Wells section, click SD1.

9. Verify that the reservoir is empty.

10. In the Reservoir section, click well RB1.

11. Verify that the plate is not warped. Warped plates can lead to incorrect probe height

adjustment.

12. Place the plate on the plate carrier with well A1 positioned as indicated on the plate

carrier.

13. Click Retract to retract the plate carrier.

14. Type a name for the plate in the Plate Name box.

15. Click Auto Adjust Height. The probe automatically adjusts itself to the locations you

selected.

NOTE: The probe height is automatically set to 0.98 mm. The probe

automatically adjusts this distance from the bottom of the plate, or
calibration disks or spheres.

16. Click Eject to eject the plate holder. If you used alignment disks or spheres, remove them

from the plate.

xPONENT for MAGPIX 4.2 Software User Manual
10