FIGURE 2.

LED Image-Based Analysis

1 Beads in chamber

2 Red LED (635 nm)

3 CCD Imager

4 Green LED (525 nm)

Running Assays with MAGPIX

General Guidelines

WARNING:

Modifying or deleting xPONENT system files can cause
degradation of system performance. Repair modified or deleted
xPONENT system files by uninstalling and re-installing the
xPONENT software. Luminex recommends that you contact
Luminex Technical Support before uninstalling and reinstalling
xPONENT.

WARNING:

Using unauthorized third-party software with xPONENT
software can result in corruption or failure of the xPONENT
software. Use third-party software at your own risk. The
operation of the system software is validated only when it runs
alone on the dedicated PC.

NOTE: If you are using a screen saver on the PC on which xPONENT is

installed, xPONENT prevents it from activating. A dialog box opens
each time xPONENT is launched, recommending that the screen
saver and any power management settings be turned off.