Plate layout subtab – Luminex xPONENT for MAGPIX 4.2 User Manual

Page 48

Advertising
background image

Assay Control Information - Lists the selected control reagents. The list displays the

Reagent, Name, Lot Number, Expiration, and Manufacturer. Existing control lot
information can be applied or new information can be typed manually.

Apply Ctrl Lot - Opens the Select Lot dialog box. Select a lot from the list and then

click OK.

Show Concentration - Expected, Low, and High set the expected, lowest, or highest

acceptable concentration of the analyte in the sample.

Apply Values - Applies a value down or across the list of analytes.

Cancel - Returns to the Batches tab.

Back - Returns to the previous tab.

Next - Continues to the Plate Layout tab.

Plate Layout Subtab

Batches > Batches > Plate Layout

Use this tab to define commands that apply to one or more wells. You can also define off-
plate and maintenance commands. This tab contains the following:

Plate Image - This is a representation of the plate. Each well is displayed as a circle on the

grid. Well commands appear in the appropriate circles as you assign them to wells on the
plate.

Command Sequence - Contains the command sequence for the active plate. The list

includes all active wells, the type of command (Unknown, Standard, Control, Background,
or assigned maintenance command), ID, and dilution factor. Double-click the ID field to
type an ID. Double-click the Dilution field to type a dilution factor.

NOTE: A command's ID and Dilution fields have a blue border around them

if they can be double-clicked to type information.

Move Command - These arrows move a selected command up or down in the Command

Sequence list, changing the acquisition order.

Batches Page

35