MAESTRO User’s Manual Revision 8.1

27

Fig. 2-12 The Wavelength menu windows

The MAESTRO automatically recognizes the latest energy and power detector for accurate auto-
calibration. More importantly, it takes advantage of our Personal wavelength correction

™: it reads the

memory in the Smart Interface connector to provide a wavelength correction based on spectral data
measured from that specific detector. Your measurements across the band have never been this precise
or easy.

You can also adjust for a wavelength other than the preset wavelength. The MAESTRO will use the pre-
programmed data in the detector’s EEPROM. When working at such a wavelength, it will not be available

in the Wavelength menu; you must manually enter the desired wavelength by touching the edit

button. A popup window menu will appear prompting you to enter the wavelength. When the wavelength

is set, touch the OK button

to save the changes, or touch the cancel button

to ignore the

changes. The MAESTRO only allows you to choose va

lues that fall within the detector’s range. If you

enter an invalid value, a menu pops up to signal the error and the MAESTRO automatically selects the
default value, which is the wavelength used for calibration at Gentec-EO during manufacturing or
subsequent service.