Set Personal Wavelength Correction
This command is used to specify the wavelength being used on the detector. The
EEPROM in the detector contains measured spectral data for a wide range of wavelengths. If the
wavelength input by the user is different from the predefined list of wavelengths in the device, a
custom value is interpolated. Specifying zero as a wavelength or providing an out-of-bound value
as a parameter restores the default settings. A valid value is set between the lowest and highest
wavelengths supported by the device, and it should not be a floating point value.
Text Command
Parameters
Return
SWA
Wavelength
ACK
Default: Calibration wavelength, (typically 1064 nm)
*SMU
Set Multiplier
This command is used to set the
multiplier’s value. Note that the MAESTRO only allows
one multiplier instead of 2 unlike the SOLO2. This means that only Multiplier 1 can be set.
Text Command
Parameters
Return
SMU
Multiplier 1 +, <Multiplier value>
ACK
The following example sets multiplier 1 to 3.3.
Example
Response from MAESTRO
Text Command:
*SMU <space> 1<space> 3.3
<enter>
ACK <CR> <LF>
*SOU
Set Offsets
This command is used to set the
offset’s value. . Note that the MAESTRO only allows
one offset instead of 2 unlike the SOLO2. This means that only Offset 1 can be set.
Text Command
Parameters
Return
SOU
Offset 1 + <Offset value>
ACK
The following example sets offset 1 to 1.5 milli.
Example
Response from MAESTRO
Text Command:
*SOU <sapce> 1 <space> 0.0015
<enter>
ACK <CR> <LF>
The other option available is the Zero-offset. The Zero-offset operation is done first,
before those of the Multipliers and Offsets