Gentec-EO MAESTRO User Manual

Page 9

Advertising
background image

MAESTRO User’s Manual Revision 8.1

8

Accuracy

3

1.0 %

50 µV < 6 kHz

2%

50 µV 6kHz to 10 kHz

Default Trigger Level

2 %

SOFTWARE
TRIGGER LEVEL

0.1% to 99.9%, 0.1% resolution

Repetition
Frequency

4

2 kHz for data acquisition in the real time full screen display, no missing

point

1.3 kHz for data acquisition in real time with time stamp, no missing point

2 kHz in statistics mode, no missing point

From 2 kHz to 10kHz the MAESTRO will send the latest energy pulse

every 500us. (corresponding to a subsampling at 2 kHz).

Frequency
Measurement

Accurate frequency measurement up to 14kHz

0-1000 Hz: 0.1% accuracy

1000-2000 Hz: 0.5%

2000-14000 Hz: 1%

Statistics

Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability,

Time, Pulse #, Repetition Rate, Avg Power.

Data Storage

Continuous on USB stick

General Specifications

Digital Display

112.9 x 84.7mm RGB color LCD with touchscreen, 640 x 480 Pixels

Display Rate

3 Hz numeric display

15 Hz bar graph & needle display

Scope

500 divisions

Data Displays

Real time, Scope, Averaging, Statistics, Digital tuning needle

User input correction
factors

1 multipliers and 1 offsets (7 digits floating point)

Analog Output

0

– 1 volt, full scale, ± .5%

Internet Upgrades

USB STICK

PC Serial Commands

USB & Ethernet

5

& RS-232

No missing point
throughput serial

300 Hz

3

Including linearity, detector dependant.

4

In a full size statistic’s windows, no scope display and no serial communication.

5

USB cable included. RS-232 cable must be purchased separately.

6

Backlight at 50% in real time display. The scope and averaging displays decrease the autonomy by

20%.

Advertising