MAESTRO User’s Manual Revision 8.1

48

2.4.5

Startup Config

The MAESTRO can be configured to store and load certain settings for a specific measurement head in
the Startup Config Menu. By default, the MAESTRO will always remember the last saved settings and
load them at startup (Autosave).

To load factory settings at startup, select Use factory settings. Note that selecting this option will not
change the current configuration, but will only load the factory settings at the next reboot.

It is also possible to save a specific configuration. To do so, change all the desired settings and return to
the Startup Config Menu. Touch the Apply button. The Use current configuration check box will be
selected and the MAESTRO’s current configure will be saved. Even if the settings are changed after this
operation, these changes will not be saved and at startup the MAESTRO will be configured in the same
manner as when the Apply button was touched.

Fig. 2-28 Startup Config menu window

These are the saved parameters:

Range;

Anticipation (on or off);

Attenuation (on or off);

Wavelength;

Multiplier and Offset;

Trigger level;

Acquisition sample rate;

Acquisition and statistics duration

Acquisition time stamp (on or off);

Save statistics to file (on or off).

Measure Mode (Power, Single Shot Energy, dBm, Energy)

Display (Scope, Needle, Statistics)

Number of Digits