Tatistics duration settings popup window – Gentec-EO MAESTRO User Manual
Page 45
![background image](/manuals/347655/45/background.png)
MAESTRO User’s Manual Revision 8.1
44
Fig. 2-26 Statistics duration settings popup window
The statistical parameters that are calculated are listed in Table 2.4.
Table 2.4. Statistical Parameters
Statistical Parameters
Power
Energy Definition
Current value
Value of the most recent measurement
Average value
Average from the start of values in the sample, E
avg
or P
avg
Maximum value
Highest value in the sample period, E
max
or P
max
Minimum value
Lowest value in the sample period, E
min
or P
min
RMS stability
Root mean square stability represents the standard deviation
as a percentage of the average.
100
avg
E
STD
RMS
,
100
avg
P
STD
RMS
PTP Stability
Shows the spread between the highest and lowest point in the
sample as a percent.
100
min
max
avg
E
E
E
PTP
,
100
min
max
avg
P
P
P
PTP