*TLA
Set Trigger Level
This command sets the internal trigger level when using the device in energy reading
mode.
Text Command
Parameters
Return
TLA
Trigger Level (percentage)
ACK
Default: 2%
The value should be set between 1 and 100 (floating point values are allowed). In text mode, you
may add a “%” symbol after the value for clarity.
Example
Response from MAESTRO
Text Command:
*TLA <space> 15.4%
<enter>
ACK <CR> <LF>
*ESU
Enable Statistics
This command is used start, stop and reset the statistics calculating process on the data currently
acquisitioned by the specified channel (upper or lower).
Text Command
Parameters
Return
ESU
0, 1 (to Stop, PLAY)
ACK
Default: stop
*SCA
Set Energy Mode
This command is used to toggle Energy mode when using a wattmeter.
Text Command
Parameters
Return
SCA
1 to turn On, 0 to turn Off
ACK