MAESTRO User’s Manual Revision 8.1
69
Zero Offset
This command subtracts the current value from all future measurements the moment the
command is issued to set a new zero point.
Text Command
Parameters
Return
SOU
None
Clear Zero Offset
This command undoes the Zero Offset command to set the zero point at zero.
Text Command
Parameters
Return
COU
None
Set diode zero.
This command subtracts the current value for all available scales from all future measurements
the moment the command is issued to set a new zero point for a photodiode only.
Text Command
Parameters
Return
SDZ
None
DONE
Set User Multipliers
This command is used to set the value of the multipliers. The default value is 1.
Text Command
Parameters
Return
MUL
8 characters numeral value
The following example sets multiplier = 33
Example
Response from MAESTRO
Text Command:
*MUL3300000
Or
*MUL3.3000e3
Get User Multiplier
This command returns the MAESTRO’s correction multiplier value.
Text Command
Parameters
Return
GUM
None
Current multiplier value