Gentec-EO MAESTRO User Manual

Page 66

Advertising
background image

MAESTRO User’s Manual Revision 8.1

65

Get Trigger Level

This command returns the trigger level in %. The value is between 0.1% and 99.9%.

Text Command

Parameters

Return

GTL

None

Returns the trigger level
in %.

Example

Response from MAESTRO

Command:

*GTL

Trigger Level : 2.0 \r\n

Get Mode Display

This command returns the MAESTRO’s power mode. Depeding on the head, it can be power
mode in W, power mode in dBm (DBM), energy mode in J, Single Shot Energy mode in J (SSE),
or simply no detector is connected to the MAESTRO.

Text Command

Parameters

Return

GMD

None

POWER = 0
ENERGY = 1
SSE = 2
DBM = 6
NO_DETECTOR = 7

Example

Response from MAESTRO

Command:

*GMD

Mode : 0 \r\n

MEASUREMENT COMMANDS -- DATA ACQUISITION

Query Current Value

This command is used to query the value that is currently

being displayed on the device’s screen.

The value is displayed in watts or in joules.

Text Command

Parameters

Return

CVU

None

Current value

For example, a 12 milliwatts reading would be displayed like this:

Example

Response from MAESTRO

Command:

*CVU

0.012 <CR> <LF>