Get Personal Wavelength
This command returns the wavelength in nm being used on the detector.
Text Command
Parameters
Return
GWL
None
Returns the wavelength
value in nm
Example
Response from MAESTRO
Command:
*GWL
PWC : 1064 \r\n
MEASUREMENT COMMANDS -- CONTROL
Set Anticipation
This command is used to enable or disable the anticipation processing when the device is
reading from a wattmeter. The anticipation is a software-reading acceleration algorithm that
provides faster readin
gs using the detector’s calibration.
Text Command
Parameters
Return
ANT
1 to turn On, 0 to turn Off
Default: On
The following example sets the anticipation.
Example
Response from MAESTRO
Command:
*ANT1
Get Anticipation Status
This command returns the anticipation status. If the anticipation is not available, it will always be
at “off”.
Text Command
Parameters
Return
GAN
None
1: On
0: Off
Example
Response from MAESTRO
Command:
*GAN
Anticipation : 0 \r\n