Gentec-EO MAESTRO User Manual

Page 84

Advertising
background image

MAESTRO User’s Manual Revision 8.1

83

*CAU

Download data

This command is used to send data to the serial port according to the data sampling

setting.

Text Command

Parameters

Return

CAU

None

Data in ASCII

*ANO

Set Analog Output

This command is used to enable or disable the output of the current value on the analog port of
the device.

Text Command

Parameters

Return

ANO

1 to Enable, 0 to Disable

ACK

*CSU

Stop the CAU Command

This command is used to stop the real time transfer enable by the CAU Command.

Text Command

Parameters

Return

CSU

None

N/A

Query New Value Ready

This command is used to check whether a new value is available from the device. Though
optional, its use is recommended when used with single pulse operations.

Text Command

Parameters

Return

NVU

Available/ Not Available <CR> <LF>

1/0

Example

Response from MAESTRO

Text Command:

*NVU <enter>

New Data Not Available <CR> <LF>

Note that the Query Current Value and Query Statistic Data commands will return the current
values from the device even if they have not been updated since the last query.