MAESTRO User’s Manual Revision 8.1

79

COMMANDS

*SSA

Set Scale

This command is used to force the display of the current data into a specific range. The

lower range is always zero, and the higher ranges can be found in the table below. The Auto
scale applies the best scale for the current values in real time. In text mode, the parameter must
be one of the identifiers in the table below. Remember that the serial protocol is not case-
sensitive in text mode.

Text Command

Parameters

Return

SSA

Scale identifier or index

ACK

Scale Identifiers:

Text Mode

Scale

Auto

optimum

1p

1 picowatt or picojoule

3p

3 picowatts or picojoules

10p

10 picowatts or picojoules

30p

30 picowatts or picojoules

100p

100 picowatts or picojoules

300p

300 picowatts or picojoules

1n

1 nanowatt or nanojoule

3n

3 nanowatts or nanojoules

10n

10 nanowatts or nanojoules

30n

30 nanowatts or nanojoules

100n

100 nanowatts or nanojoules

300n

300 nanowatts or nanojoules

1u

1 microwatt or microjoule

3u

3 microwatts or microjoules

10u

10 microwatts or microjoules

30u

30 microwatts or microjoules

100u

100 microwatts or microjoules

300u

300 microwatts or microjoules

1m

1 milliwatt or millijoule