Ata sampling parameters window menu – Gentec-EO MAESTRO User Manual
Page 47
MAESTRO User’s Manual Revision 8.1
46
Fig. 2-27 Data sampling parameters window menu.
The different available options are presented in Table 2.5. To edit one of the data sampling parameters,
touch the edit
button. Touch the OK button
to save the changes or touch the cancel button
to ignore the changes.
Table 2.5 Data Acquisition Parameters
PARAMETER
Choices
Description
Default
Sample Rate
Power only
Integers
pts/Second
pts/Minute
pts/Hour
pts/Day
Current setting is 10 samples per seconds.
Sets the time between each sample. Specify the
number of points per unit of time. [for example,
for 1 second between samples, set to 60 points
per minute].
Next, you must set the time period for the
number of points entered.
Maximum is 10 points/second,
Minimum is 1 point/day.
Time between samples = 1/(sample rate).
10 per sec
Sample Rate
Energy only
Integers
1/x pulses
Sets the fraction (1/x pulses) of the incoming
pulses sampled for the statistics calculations and
data recording.
1/1 pulses of the
incoming pulses
Duration
Time format
The time period for which samples are reported
(to the display and output). The format of the
duration is dd: hh:mm:ss, which correspond to
1 minute