Ata sampling parameters window menu – Gentec-EO MAESTRO User Manual

Page 47

Advertising
background image

MAESTRO User’s Manual Revision 8.1

46

Fig. 2-27 Data sampling parameters window menu.

The different available options are presented in Table 2.5. To edit one of the data sampling parameters,

touch the edit

button. Touch the OK button

to save the changes or touch the cancel button

to ignore the changes.

Table 2.5 Data Acquisition Parameters

PARAMETER

Choices

Description

Default

Sample Rate

Power only

Integers

pts/Second
pts/Minute
pts/Hour
pts/Day


Current setting is 10 samples per seconds.

Sets the time between each sample. Specify the
number of points per unit of time. [for example,
for 1 second between samples, set to 60 points
per minute].
Next, you must set the time period for the
number of points entered.
Maximum is 10 points/second,
Minimum is 1 point/day.
Time between samples = 1/(sample rate).

10 per sec

Sample Rate

Energy only

Integers

1/x pulses

Sets the fraction (1/x pulses) of the incoming
pulses sampled for the statistics calculations and
data recording.

1/1 pulses of the
incoming pulses

Duration

Time format

The time period for which samples are reported
(to the display and output). The format of the
duration is dd: hh:mm:ss, which correspond to

1 minute