Gentec-EO MAESTRO User Manual

Page 65

Advertising
background image

MAESTRO User’s Manual Revision 8.1

64

Example

Response from MAESTRO

Command:

*GAS

AutoScale : 1 \r\n

Display Valid Range

This command is used to display all of the valid ranges the connected head supports. The
ranges are displayed in scale index. Please refer to the Set Range section for the table
correspondence.

Command

Parameters

Return

DVS

None

The valid scale index.

The following example is for a UP19K connected to the MAESTRO, which can have the following
ranges:

100 mW;

300 mW;

1 W;

3 W;

10 W;

30 W.

Example

Response from MAESTRO

Command:

*DVS

[22] : 100.0 m <CR> <LF>

[23] : 300 m <CR> <LF>

[24] : 1.000 <CR> <LF>

[25] : 3.00 <CR> <LF>

[26] : 10.00 <CR> <LF>

[27] : 30.0 <CR> <LF>

Set Trigger Level

This command sets the internal trigger level when using the device in energy reading mode.

Text Command

Parameters

Return

STL

Trigger Level (in percentage) must
be 4 numeral values

Default: 2%

The value should be set between 0.1 and 99.9.

Example

Response from MAESTRO

Text Command:

*STL15.4 (15.4%)

*STL0.20 (.2%)

Advertising