Figure b max. and min. humidity set ranges – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 100

Advertising
background image

IM 11M12A01-05E

8-3

8. Detailed Data Setting

Humidity setting range
The minimum humidity is set to 0% H

2

O or ranges from 26 to 100% H

2

O. The maxi-

mum humidity ranges from 25% to 100% H

2

O, and must be greater than 0.8 times plus

23 the humidity set for the minimum.

Setting example 1
If the setting (for a 4 mA current) is 0% H

2

O, you must set the maximum (20 mA) point

at more than 25% H

2

O.

Setting example 2
If the setting (for a 4 mA current) is 26% H

2

O, you must set the maximum (20 mA)

point at more than 44% H

2

O, (26 3 0.8 + 23% H

2

O). (Numbers after the decimal point

are rounded up.)

5

15

25

35

45

55

65

75

85

95

5

15

25

35

45

55

65

75

F8.1.1.2.EPS

Minimum humidity (for a 4-mA current), % H

2

O

Maximum humidity (for a 20-mA

current), % H

2

O

Ranges over which oxygen
concentrations can be set

Outside ranges

Figure B Max. and Min. Humidity Set Ranges