B1 0, Span y, Open/20.84 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 104: Zero y, Open/0.89, Calend

Advertising
background image

IM 11M12A01-05E

8-7

8. Detailed Data Setting

F8.1E.EPS

Switch operations

Display

Output hold time
during calibration

>

ENT

b

1

0

>

ENT

CAL

>

ENT

CAL

>

ENT

SPAn

Y

>

ENT

21.00

%

>

ENT

OPEn/20.84

>

ENT

20.84

%

>

ENT

ZEro

Y

>

ENT

0.98

%

>

ENT

OPEn/0.89

>

ENT

0.89

%

>

ENT

CALEnd

>

ENT

b

10

>

ENT

Measured-value display

>

>

>

>

>

>

>

>

>

>

>

>

>

>

Figure 8.1 Definition of during calibration
During semi-automatic calibration, “under calibration” is the time, starting when a
calibration instruction is executed with an infrared switch or a contact input, to make a
series of calibrations, until the preset output stabilization time elapses.

During automatic calibration, “under calibration” is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization
time elapses.

(4) “Error” appears when Error 1 to Error 4 are being issued