7-21

7. Startup

Table 7.13 Calibration Setup Procedure

T7.13E.EPS

Switch operation

Display

Description

Display after the password has been entered.

Set the zero-gas concentration. Switch the parameter code to B01.
Here, set 0.98%.

Touch the [ENT] key to display the currently set value.

Touch the [>] key to move the position of the digit that is flashing to 1.

Touch the [ ] key to change to 0.

Touch the [>] key to move the position of the digit that
is flashing to the right one digit.

Touch the [ ] key to change the numeric value to 9.

Touch the [>] key to move the position of the digit that
is flashing to the right one digit.

Touch the [ ] key to change the numeric value to 8.

Touch the [ENT] key to have all the digits flash.

Touch the [ENT] key again to stop the flashing.

Touch the [ENT] key once again to switch to the
parameter code selection display.

A

01

b

01

0

01.00

%

00

1

.00

%

00

0

.00

%

000.

0

0

%

000.

9

0

%

000.9

0

%

000.9

8

%

000.98

%

000.98

%

b

01

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

b0

3

0

0

0

b

03

Basic
panel
display

Next, set the calibration mode. Switch the parameter code to B03.

Touch the [ENT] key to display the currently set value.
If it is 0, you can leave it as is. If it is other than 0, change it to 0 (zero).

Touch the [ENT] key. The numeric value will flash.

Touch the [ENT] key again to stop the flashing.

Touch the [ENT] key once again to switch to the parameter code selection display.

Touch the [>] key together with the [ENT] key to return to the basic panel display.
(This is not required if you proceed to make another setting.)
(The displayed numeric characters indicate the measurement gas concentration.)

>

>

>

>

>

>

>

>

>

>

>

>

>

>

>

>

>

>

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

.

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

.

.

.

Set the span-gas concentration by above procedure, set 21 % to B02.