IM 11M12A01-05E

5-4

5.2 Wiring for Analog Output

This wiring is for transmitting 4 to 20 mA DC output signals to a device, e.g. recorder.
Maintain the load resistance including the wiring resistance of 550V or less.

+

-

AO(+)
AO(

-

)

FG

F5.4E.EPS

Shielded cables

Analyzer

Receiver

Figure 5.4 Wiring for Analog Output

5.2.1

Cable Specifications

Use a 2-core shielded cable for wiring.

5.2.2

Wiring Procedure

(1) M4 screws are used for the terminals. Use crimp-on terminals appropriate for M4

terminal screws for cable connections. Ensure that the cable shield is connected to
the FG terminal of the equipment.

(2) Be sure to connect (+) and (2) polarities correctly.

CAUTION

• Before opening the detector cover, loosen the lock screw. If the screw is not loosened

first, the screw will damage the cover and the terminal box will require replacement.
When opening and closing the cover, remove any sand particles or dust to avoid
gouging the thread.

• After screwing the cover on the equipment body, secure it with the lock screw.