14 output-related items in group c – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 141

Advertising
background image

IM 11M12A01-05E

10-14

Output-related Items in Group C

Code

Item

Tuning

Engineering Default setting
unit

C01

Analog output

0: Oxygen concentration

Humidity

1: Humidity

2: Mixed ratio

C03

Output mode

0: Linear

Linear

1: Logarithm

C04

Output during warm-up 0: Held at 4 mA
up

1: Held at 20 mA

2: Set value remains held.

Held at 4 mA.

C05

Output during

0: Not held

maintenance

1: Held output just before

Held output just before

maintenance service.

maintenance service.

2: Set value remains held.

C06

Output during

0: Not held

Held output just before

calibration 1: Held output just before

calibration.

calibration.

2: Set value remains held.

C07

Output in abnormal 0: Not held

Held output at a preset

state

1: Held output just before

value.

abnormal state occurs.

2: Set value remains held.

C11

Min. oxygen

See Section 8.1.

% O

2

0% O

2

concentration

C12

Max. oxygen

See Section 8.1.

% O

2

25% O

2

concentration

C13

Minimum humidity

See Section 8.1.

% H

2

O 0% H

2

O

C14

Maximum humidity

See Section 8.1.

% H

2

O 25% H

2

O

C15

Minimum mixing ratio See Section 8.1. kg/kg

0 kg/kg

C16

Maximum mixing ratio See Section 8.1. kg/kg

0.2kg/kg

C30

Output damping

0 to 255

Seconds

0 second

factor

C31

Set value during

2.4 to 21.6

mA

4 mA

warm-up

C32

Set value during

2.4 to 21.6

mA

4 mA

maintenance

C33

Set value during

2.4 to 21.6

mA

4 mA

calibration

C34

Set value in abnormal 2.4 to 21.6

mA

3.4 mA

state

T10.6C.EPS

Advertising