12 current output, 13 response time, 14 internal resistance of cell – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 132

Advertising
background image

IM 11M12A01-05E

10-5

10. Other Functions

10.1.12 Current Output

The analog output current is displayed.

10.1.13 Response Time

The cell’s response time is obtained in the procedure shown in Figure 10.1. If only
either a zero-point or span calibration has been carried out, the response time will not be
measured just as it will not be measured in manual calibration.

mA

90%

100%

The response time is obtained after the corrected calibration curve has been found. The response time
is calculated, starting at the point corresponding to 10% of the analog output up to the point at 90% of
the analog output span. That is, this response time is a 10 to 90% response.

10% of analog

output span

Five minutes maximum

Response time

Start calibration

Time

Calibration
complete

F10.1E.EPS

Figure 10.1 Functional Drawing of Response Time

10.1.14 Internal Resistance of Cell

A new cell (sensor) indicates its internal resistance of 200 V maximum. As the cell
degrades, so will the cell’s internal resistance increase. The degradation of the cell
cannot be found only by changes in cell’s internal resistance, however. Those changes
in the cell’s internal resistance will be a hint to knowing the sensor is degrading. The
updated values obtained during the calibration are displayed.