3 checking input contacts, Setting -1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 92

Advertising
background image

IM 11M12A01-05E

7-19

7. Startup

7.9.3

Checking Input Contacts

Follow Table 7.11 to check the input contacts. The table uses an example with input
contact 1.

Table 7.11 Checking Input Contacts

T7.11E.EPS

>

>

>

>

>

>

.

.

..

.

.

ENT

ENT

ENT

ENT

ENT

ENT

Switch operation

Display

Description

A

01

G

01

G

0

1

G

2

1

0

G

21

The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate flashing.

Display after the password has been entered.

Touch the [ ] key to switch to Group G.

Touch the [>] key to move the position of the digit
that is flashing to the right one digit.

Touch the [ ] key to enter 2.

Touch the [ENT] key. 0 is displayed with the contact open.
If the contact is closed, the display will be 1 (one).
This enables you to check whether or not the wiring
connections have been properly made or not.

Touch the [>] key together with the [ENT] key to return to
the basic panel display.

.

.