2. Specifications

2.4.4 Zero-gas Cylinder (part number: G7001ZC)

The gas from this cylinder is used as the calibration zero gas and detector purge gas.

Standard Specifications
Capacity : 3.4 l

Filled pressure : 9.8 to 12 MPa G

Composition : 0.95 to 1.0 vol% O

2

in N

2

(Note) Export of such high pressure filled gas cylinders to most countries is prohibited

or restricted.

485

325

[140

Unit : mm

F2213.EPS

Weight : Approx. 6 kg

2.4.5 Pressure Regulator for Gas Cylinder (part number: G7013XF or G7014XF)

This regulator valve is used with the zero gas cylinders.

Standard Specifications
Primary Pressure: Max. 14.8 MPa G

Secondary Pressure: 0 to 0.4 MPa G

Connection : Inlet W22 14 threads, right hand screw

Outlet Rc1/4 or 1/4FNPT

Material: Brass body

´

J

HI

H

O

C

A

K

A

T

Regulator handle

W22 (Right-

handed screw)

Stop valve

Approx. 174

Approx.1

12

IN

HI

H

AC

A

K

T

´ J

O

Outlet

p

Part No.

G7013XF

G7014XF

p Outlet

Rc1/4

1/4 NPT female screw

Secondary

pressure gauge

Primary

pressure gauge

Secondary

safety valve

Primary

safety valve

Approx. 82

Approx. 59

Approx. 163

Unit : mm