3.4 Installation of the Calibration-gas Unit Case (E7044KF)

The calibration gas unit case is used to store the G7001ZC zero gas cylinders.

3.4.1 Location

The following should be taken into consideration:

(1) Easy access for cylinder replacement
(2) Easy access for checking
(3) Near to the detector and converter as well as the flow setting unit.
(4) The temperature of the case does not exceed 408 C due to rays of the sun or radiated

heat.

(5) No vibration

3.4.2 Mounting

Mount the calibration gas unit case on a pipe (nominal JIS 50A) as follows:

(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A : O.D. 60.5 mm) for

mounting the flow setting unit. (The combination of the calibration gas unit case and
the calibration gas cylinder weighs approximately 4.2 kg.)

(2) Mount the unit case on the pipe by tightening the nuts with the U-bolt so that the

metal fitting is firmly attached to the pipe.

F3.16E.EPS

A pipe to be mounted
(nominal JIS 50A: O.D. 60.5 mm)

Figure 3.7 Pipe Mounting