7. Startup

7.4.2

Display Configuration

The parameter codes provided for the equipment are used to control the equipment
display panels (see below). By selecting appropriate parameter codes, you can conduct
calibration and set operation parameters. Figure 7.4 shows the configuration of display
items. The parameter codes are listed in groups of seven; which are briefly described in
Table 7.2.

To enter parameters, you first need to enter the password.

Touch the [ >] key and [ ENT ] key at same time to revert to the main screen.

F7.4E.EPS

Group G setup display

Basic panel display

Password entry display

Group A setup display

Group B setup display

Group C setup display

Parameter code
selection display

Group D setup display

Group E setup display

Group F setup display

%

Figure 7.4 Display Configuration

Table 7.2 Display Functions

Basic panel

Displays the oxygen concentration in normal operation, or displays the detector
heater temperature while warming up. If an error or alarm arises,
the corresponding error or alarm number appears.

Password entry

Enters the password for the parameter code selection display.

Group A setup

Displays detailed data, such as the cell voltage or temperature.

Group B setup

Sets and performs calibration and blowback.

Group C setup

Sets analog output.

Group D setup

Sets an alarm.

Group E setup

Sets the input and output contacts.

Group F setup

Selects the type of equipment and sets the parameters for computation.

Group G setup

Performs the current-loop or contact checks.