5 insulation resistance test, Insulation resistance test -9 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual

Page 49

Advertising
background image

IM 11M12A01-05E

3-9

3. Installation

3.5 Insulation Resistance Test

Even if the testing voltage is not so great it causes dielectric breakdown, testing may

cause deterioration in insulation and a possible safety hazard. Therefore, conduct this
test only when it is necessary.

The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied
for as short a time as practicable to conform that insulation resistance is 20 MV or more.

Remove wiring from the converter and the detector.

1. Remove the jumper plate located between terminal G and the protective grounding

terminal.

2. Connect crossover wiring between L and N.
3. Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to

the crossover wiring, and (-) terminal to ground.

4. Turn the insulation resistance tester ON and measure the insulation resistance.
5. After testing, remove the tester and connect a 100 kV resistance between the cross-

over wiring and ground to discharge.

6. Testing between the heater terminal and ground, contact output terminal and ground,

analog output/input terminal and ground can be conducted in the same manner.

7. Although contact input terminals are isolated, insulation resistance test cannot be

conducted because the breakdown voltage of the surge-preventing arrester between
the terminal and ground is low.

8. After conducting all the tests, replace the jumper plate as it was.

1

DI-1

2

DI-2

3

DI-C

4

DO-1

5

DO-1

6

DO-2

7

DO-2

8

FG

9

AO

(+)

10

AO

(-)

11

L

12

N

13

G

14

FG

F3.17E.EPS

Remove
jumper
plate

Insulation

resistance

tester

Insulation

resistance

tester

Crossover wiring

Crossover wiring

Contact input 1

Contact input 2

+

-

+

-

Insulation

resistance

tester

+

-

Advertising