5. Wiring

5.3 Wiring Power and Ground Terminals

Wiring for supplying power to the analyzer and grounding the equipment.

1

2

C

DO

DO

DI

1

2

FG

L

N

G

FG

+

-

AO

~

100~240 V AC
50/60 Hz

F5.5E.EPS

Jumper plate

Grounding to the ground terminal
on the equipment case

Equipment case

Lock washer

Crimp contact of
the grounding line

Grounding
terminal

Ground

Figure 5.5 Power and Grounding Wiring

5.3.1

Wiring for Power Line

Connect the power wiring to the L and N terminals of the equipment. For a three-core
cable, ground one core appropriately. Proceed as follows:

(1) Use a two-core or three-core shielded cable.
(2) M4 screws are used for the terminals. Use crimp-on terminals appropriate for M4

terminal screws for cable connections.

5.3.2

Wiring for Ground Terminals

The ground wiring of the analyzer should be connected to either the ground terminal of
the equipment case or the terminal inside of the equipment. Proceed as follows:

(1) Keep the ground resistance of 100V or less (JIS Class D grounding).
(2) When connecting the ground wiring to the ground terminal of the equipment case, be

sure that the lock washer is in contact with the case surface (see Figure 5.5.).

(3) Ensure that the jumper plate is connected between the G terminal and the FG

terminal of the equipment.

(4) The size of external ground screw thread is M4. Each cable should be terminated

corresponding crimp-on terminals.